Researchers evaluating calibration strategies for HyperSight cone-beam computed tomography have demonstrated that adding solid-water extensions or restricting blade collimation substantially improves dosimetric agreement with simulation computed tomography when using a standard thin CIRS Model 062 M phantom. The methodological investigation compared three practical setup approaches on an Ethos HyperSight platform, revealing that direct full-field scanning introduces severe dosimetric discrepancies that can compromise dose calculations in adaptive radiotherapy.
HyperSight cone-beam computed tomography equipped with Acuros reconstruction is designed to facilitate direct dose calculation during adaptive radiotherapy workflows. However, the calculation accuracy achieved by this imaging modality relies heavily on a dependable calibration curve that correlates computed tomography numbers with electron and material density values.
To evaluate these calibration dynamics, the study analyzed three distinct acquisition protocols using a CIRS Model 062 M phantom. The phantom was initially scanned on a dedicated simulation computed tomography scanner to establish a reference curve linking computed tomography numbers to electron and material density. The identical phantom was subsequently imaged on an Ethos HyperSight cone-beam computed tomography system under three experimental configurations designated as Method A, Method B, and Method C.
The three acquisition techniques tested specific mechanical and phantom adjustments during the calibration process. Method A utilized 4-centimeter blade collimation oriented along the y-direction. Method B employed an unextended 24.2-centimeter y-direction blade collimation across the full field of the phantom.
The clinical validity of each resulting calibration curve was subsequently tested using anthropomorphic phantoms representing three anatomical regions: the head, the thorax, and the pelvis. Each anthropomorphic phantom was scanned on both the reference simulation computed tomography scanner and the Ethos HyperSight platform. Patient-derived radiotherapy plans, comprising volumetric modulated arc therapy and intensity modulated radiotherapy representing the three anatomical sites, were transferred onto the corresponding phantom image datasets to calculate radiation doses.
Dose calculations derived from the three experimental HyperSight calibration curves were compared against the reference simulation computed tomography plans using dose-volume metrics and three-dimensional global gamma analysis. The researchers evaluated gamma agreement across several strict distance and dose criteria, specifically 1 millimeter with 1 percent, 1 millimeter with 2 percent, 2 millimeters with 2 percent, 2 millimeters with 3 percent, and 3 millimeters with 3 percent, applying a 10 percent low-dose evaluation threshold.
Visual inspection of the reconstructed images revealed that direct 24.2-centimeter blade scanning under Method B generated prominent streak artifacts. In contrast, 4-centimeter blade collimation under Method A and the solid-water-extended setup under Method C produced overall image quality that remained comparable to the reference simulation computed tomography scans.
Quantitative three-dimensional gamma evaluations demonstrated pronounced performance differences among the three calibration techniques. Under the rigorous 1 millimeter with 1 percent evaluation criteria, Method A produced body pass rates of 96.6 percent for the head, 99.5 percent for the thorax, and 99.9 percent for the pelvis, with corresponding planning target volume pass rates of 99.0 percent, 85.7 percent, and 99.9 percent. In contrast, Method B exhibited marked degradation in planning target volume pass rates, dropping to 40.4 percent in the head, 58.9 percent in the thorax, and 7.0 percent in the pelvis. Method C fully restored agreement, achieving body pass rates between 99.8 percent and 100.0 percent alongside planning target volume pass rates ranging from 98.2 percent to 99.9 percent. When evaluated at the 2 millimeters with 2 percent threshold, all approaches except Method B exceeded a 99 percent pass rate.
The study authors concluded that performing full-field cone-beam computed tomography calibration without solid-water longitudinal extension sharply reduces dosimetric agreement when utilizing a thin CIRS Model 062 M phantom. Adding solid-water extension slabs provides the closest agreement with reference simulation computed tomography datasets, while narrow 4-centimeter blade collimation represents an effective and practical operational alternative whenever solid-water extension materials are unavailable in clinical settings.
Subject of Research: HyperSight cone-beam computed tomography calibration methods
Article Title: Establishing CT-number-to-electron/density curves for hypersight CBCT with a standard cirs 062 M phantom: a methodological comparison of three practical approaches
Article References: Wang, Z., Wang, X., Wu, X., Wang, X., Gu, J., Gao, H., Jiang, P., & Zhang, Y. (2026). Establishing CT-number-to-electron/density curves for hypersight CBCT with a standard cirs 062 M phantom: a methodological comparison of three practical approaches. Radiation Oncology. https://doi.org/10.1186/s13014-026-02938-2
Image Credits: AI Generated
DOI: 10.1186/s13014-026-02938-2
Keywords: HyperSight CBCT, cone-beam computed tomography, Acuros reconstruction, adaptive radiotherapy, CT-number calibration, electron density curve, material density curve, CIRS Model 062 M phantom, simulation CT, Ethos system, blade collimation, solid-water slabs, anthropomorphic phantoms, volumetric modulated arc therapy, intensity modulated radiotherapy, 3D gamma analysis, dose-volume metrics, streak artifacts
News Source: Nathaniel Bowman. (November 24, 2026). Solid-Water Extension Improves HyperSight Cone-Beam CT Calibration in Adaptive Radiotherapy. Scienmag.