In complex endoluminal procedures such as endoscopic submucosal dissection, practitioners must frequently operate multiple electrosurgical and auxiliary devices through an array of pedals while sustaining unbroken visual concentration on subtle tissue dynamics. The experimental study, conducted with eighteen practicing endoscopists from thirteen medical institutions, revealed that integrating an action camera pointed downward at the foot-pedal assembly reduced the average time required to initiate a commanded device function by 90.4 milliseconds compared to conventional operation without visual pedal feedback. Beyond accelerating response speed, the camera-assisted interface curtailed the incidence of incorrect pedal depressions and reduced head movements away from the main procedural monitor, addressing a recognized source of procedural disruption and physical fatigue in advanced digestive endoscopy.
Operating physicians must maintain an exceptionally stable endoscopic view within confined luminal spaces while simultaneously manipulating delicate instruments through the endoscope working channel. When visualization degrades, the endoscopist is often forced to reposition the endoscope and laboriously reconstruct the mucosal view, a complication that substantially lengthens total procedure time, particularly for early-career practitioners who are still consolidating the necessary psychomotor skills.
The procedural complexity of advanced therapeutic endoscopy has intensified with recent methodological and technological innovations, leading to a marked proliferation of foot-operated peripheral equipment. In emerging approaches such as underwater endoscopic submucosal dissection utilizing the water-pressure method, maintaining clear visualization requires rapid, precisely timed coordination between high-frequency cutting currents, coagulation currents, mucosal injection systems, and water-jet irrigation. To execute these disparate functions, operators must manage four or more distinct pedals located at floor level while concurrently maneuvering scope control wheels and directing auxiliary instrumentation. Because modern endoscopy demands continuous integration of scope manipulation, device articulation, and auxiliary equipment activation, sub-optimal equipment positioning and visual disconnection increase procedural friction and exacerbate the risk of occupational endoscopy-related musculoskeletal disorders.
To overcome these ergonomics and operational hurdles, the investigators designed an auxiliary monitoring framework that situates a compact action camera above the floor pedal array, routing a live video feed directly into a corner of the primary endoscopic monitor. The researchers hypothesized that providing a real-time visual anchor on the operative screen would enable clinicians to reconfirm pedal orientation without ever breaking eye contact with the therapeutic target. Crucially, the research team recognized that any decrease in gaze shifts or reduction in pedal selection mistakes would offer negligible clinical utility if achieved by forcing operators into a hesitant, overly cautious posture. Consequently, the team established participant-level mean reaction time as the primary outcome measure, explicitly testing whether visual assistance could accelerate pedal execution rather than merely altering operator caution.
The study was conducted as a rigorous, single-center simulation trial under a protocol formally reviewed and approved by the institutional ethics committee under Approval Number 202602-02 and registered with the University Hospital Medical Information Network Clinical Trials Registry under identifier UMIN000060763. To establish a robust sample size, the researchers based their statistical power calculations on a preliminary pilot evaluation involving seven expert endoscopists, wherein camera assistance achieved a mean reaction time reduction of 144.29 milliseconds with an observed effect size of Cohen’s dz equal to 1.415. Recognizing that the small pilot sample carried a substantial risk of effect size overestimation, the investigators consulted a biostatistician to model four conservative enrollment scenarios assuming twenty percent or thirty percent attenuation of the effect size at eighty or ninety percent statistical power. Selecting the most rigorous scenario—which assumed a thirty percent attenuation yielding Cohen’s dz equal to 0.99 and ninety percent power at a two-sided significance alpha of 0.05—indicated a minimum requirement of fourteen subjects. To safeguard the trial against potential data loss, recording failures, or unexpected exclusions, the investigators recruited a final cohort of eighteen endoscopists.
The experimental setup was carefully constructed to simulate the visual and physical ergonomics of an active therapeutic endoscopy suite. All participating endoscopists executed the tasks in a standing posture, with monitor height, operator-to-screen distance, and ambient room illumination held strictly uniform across trials. The physical pedal array comprised four industrial foot switches, each measuring 100 millimeters by 60 millimeters, firmly anchored to the floor with standardized spacing between adjacent units. The task interface was generated by a proprietary software application operating on a personal computer and displayed on a commercial Olympus EVIS X1 endoscopic monitor. In the camera-assisted condition, the action camera feed capturing the floor assembly was routed to the lower-left corner of the EVIS X1 monitor in a standardized picture-in-picture layout, mirroring the configuration utilized in exploratory clinical applications. In the control non-camera condition, this lower-left visual quadrant remained absent, requiring participants to rely either on tactile proprioception or direct downward visual glances.
During each simulation trial, the custom software presented one of four randomized Japanese command words corresponding to standard clinical foot functions: cut, coagulation, water jet, or injection. Reaction time was automatically registered from the millisecond the text prompt appeared until the application detected mechanical activation of the correct switch. If a participant inadvertently stepped on the wrong pedal, the application registered an incorrect pedal press and kept the timer running continuously until the participant relocated and depressed the correct pedal. Participants completed three consecutive sets of twenty trials under both experimental conditions in a randomized sequence, yielding sixty discrete trials per condition and one hundred twenty trials per participant. To reflect real-world clinical diversity, the researchers placed no restrictions on foot utilization, permitting participants to operate pedals using their dominant foot, non-dominant foot, or both feet simultaneously in alignment with their routine endoscopic habits.
To prevent motor learning, muscle memory, and spatial habituation from distorting the results, the experimental team physically rearranged the sequence of the four pedals in a randomized order prior to the start of every twenty-trial set. This mandatory restructuring applied equally across both camera-assisted and non-camera trials, ensuring that every participant continually encountered unfamiliar spatial arrangements that necessitated active positional verification before pedal activation. Endoscopists were directed first and foremost to avoid depressing the wrong pedal, and secondarily to execute the correct command as rapidly as possible while maintaining that absolute priority.
The eighteen enrolled endoscopists represented thirteen distinct medical institutions, providing institutional breadth across varied training backgrounds. Thirteen participants were male, representing 72.2 percent of the cohort, while five were female, representing 27.8 percent. The median post-graduate endoscopic experience of the cohort stood at 4 years, with a total range spanning from 2 to 24 years. Categorical stratification of professional experience revealed that twelve operators, or 66.7 percent, had practiced endoscopy for 1 to 5 years; three participants had 6 to 10 years of experience; one clinician had between 11 and 15 years; and two participants possessed 16 or more years of experience. All eighteen clinicians successfully completed all one hundred twenty simulation trials without any protocol deviations, technical data interruptions, or sample attrition, allowing the complete cohort to be included in the primary and secondary statistical evaluations.
Analysis of the primary endpoint demonstrated that camera assistance produced a statistically significant reduction in mean reaction times across the participant cohort. Under conventional non-camera conditions, the mean participant reaction time was 1280.8 milliseconds with a standard deviation of 244.0 milliseconds, whereas camera assistance lowered the mean reaction time to 1190.4 milliseconds with a standard deviation of 209.3 milliseconds. This yielded a statistically significant mean paired reduction of 90.4 milliseconds, supported by a 95 percent confidence interval ranging from 36.8 to 144.0 milliseconds and a paired t-test p-value of 0.0024. The corresponding standardized effect size was calculated at Cohen’s dz equal to 0.8394. Furthermore, exploratory correlation analysis indicated that the magnitude of reaction time improvement was not significantly correlated with the operator’s career longevity, displaying a Spearman’s rank correlation coefficient of minus 0.069 with a non-significant p-value of 0.785, suggesting that the visual overlay offered operational advantages across varying tiers of clinician experience.
Secondary outcome measures demonstrated concordant, pronounced improvements when operators utilized the camera-assisted configuration. The mean of participant-level median reaction times fell from 1207.1 milliseconds without assistance to 1141.0 milliseconds with assistance, reflecting an average within-participant median reduction of 66.1 milliseconds with a 95 percent confidence interval of 27.5 to 104.8 milliseconds and a p-value of 0.0022. Erroneous pedal presses, assessed via the Wilcoxon signed-rank test across the sixty-trial evaluation blocks, showed a statistically significant decrease from a median of 1 press with an interquartile range of 0 to 2 under conventional conditions down to a median of 0 presses with an interquartile range of 0 to 1 with camera assistance, achieving a p-value of 0.0378. The most dramatic shift appeared in operator gaze metrics evaluated from synchronized lateral video recordings: downward head shifts toward the floor plummeted from a median of 59.5 glances with an interquartile range of 45.5 to 60.0 under standard conditions to a median of 0 glances with an interquartile range of 0 to 0.75 under camera assistance, representing a highly significant behavioral shift with a p-value below 0.001.
While these quantitative metrics demonstrate clear performance enhancements in a laboratory environment, the authors highlighted several methodological limitations that temper direct extrapolation to routine surgical care. The study was conducted at a single academic center using an isolated simulation task where commands were displayed as text prompts rather than emerging organically from procedural tissue feedback. In live endoscopic operations, clinicians must integrate pedal commands while actively steering the endoscope, applying delicate tension, and managing unforeseen anatomical challenges such as arterial bleeding, severe submucosal fibrosis, respiratory excursion, and verbal coordination with surgical assistants. Furthermore, the deliberate physical randomization of foot pedals between experimental sets deliberately removed spatial familiarity to test visual verification; in standard operating theaters, pedals frequently occupy relatively fixed positions, although they remain vulnerable to intraoperative drift. Finally, the authors noted that while the lower-left picture-in-picture window did not obscure the live endoscopic field, it partially overlapped the area reserved for reviewing frozen reference images on the Olympus monitor, representing an interface trade-off that warrants dedicated clinical validation in real-world operating rooms.
Subject of Research: Endoscopic surgery foot-pedal monitoring camera simulation
Article Title: Foot‐Pedal Monitoring Camera Improves Pedal Operation During Simulated Therapeutic Endoscopy
Article References: Nakajima, R., Sano, W., Shimizu, S., Hirata, D., Iwatate, M., Fujita, M., & Sano, Y. (2027). Foot‐Pedal Monitoring Camera Improves Pedal Operation During Simulated Therapeutic Endoscopy. DEN Open, 7(1), Article e70452. https://doi.org/10.1002/deo2.70452
Image Credits: AI Generated
DOI: 10.1002/deo2.70452
Keywords: therapeutic endoscopy, endoscopic submucosal dissection, foot pedal monitoring, simulation study, reaction time, ergonomics, picture in picture display, gaze shift reduction, electrosurgical control, underwater ESD, procedural visualization, clinical simulation, surgical safety, paired t test, Olympus EVIS X1, medical device interface, human factors engineering
News Source: Ophelia Keating. (November 24, 2026). Foot-Pedal Camera System Accelerates Response Times in Simulated Endoscopic Surgery. Scienmag.